model |
LHH series medicine test box |
LHH series general medicine stability test box |
||||
LHH-150SD |
LHH-250SD |
LHH-150SDP |
LHH-250SDP |
LHH-150GSP |
LHH-250GSP |
|
Electrical requirements |
AC 220V 50HZ |
|||||
Temperature range |
0C-65℃ |
NO lighting:0C-65℃
lighting:10C-65℃ |
||||
Temperature control |
±0.5℃ |
|||||
Humidity range |
40-95%RH |
|||||
Humidity deviation |
±3%RH |
|||||
Illumination range |
N/A |
0-6000LX |
||||
controller |
Special temperature and humidity digital display instrument |
Imported touch-screen programmable controller |
||||
Interior dimension(mm) |
500*505*600 |
550*520*900 |
500*505*600 |
550*520*900 |
500*505*705 |
550*520*1020 |
Overall dimension(mm) |
630*705*1150 |
680*720*1540 |
630*705*1150 |
680*720*1540 |
630*705*1270 |
680*720*1660 |
Timing range |
3(pcs) |
Satisfying the ICH2003 Q1A(2)guiding principle /GMP2005 edition Chinese pharmacopoeia stability test conditions:
1 short-time test:40C±2C/75%RH±5%RH
2intermediate qualification:30c±2C/65%RH±5%RH
3 long-run test:25C±2C/60%RH±5%RH or 30C±2C/65%RH±5%RH
Execution standard:2005 edition pharmacopoeia medicine stability test guideline and relative items in GB 10586-2006